Abstract: Objective Ring artifacts often appear in flat-panel detector-based CT images due to the malfunction ormis-calibration of the detector elements that result in stripe artifacts in the line integral projection (sinogram) data. The ringartifacts lower the image quality and affect image-based diagnoses. Here we propose a ring artifacts removal approach basedon wavelet filtering in the sinogram domain. The line integral projection (sinogram) dataset were divided into 4 sub-sinogramdataset, and for each of them the wavelet decomposition operation was employed to produce the associated wavelet dataset,followed by filtering operation on the vertical detail band and the low-pass detail band. Wavelet reconstruction operation wasthen performed, and the weighted moving average filter was used to yield the filtered sinogram, which was processed usingfiltered back-projection (FBP) for image reconstruction. The results showed that the proposed approach could effectivelyremove the ring artifacts while preserving the structural information of the image.