Journal of Southern Medical University ›› 2014, Vol. 34 ›› Issue (06): 901-.
Previous Articles Next Articles
Online:
Published:
Abstract: Objective To evaluation the performance of a heart-type fatty acid-binding protein (H-FABP) ELISA detection kit inthe clinical diagnosis of acute coronary syndrome (ACS). Methods Plasma or serum samples from 160 suspected ACS patientshospitalized in General Hospital of PLA were examined using Lanzhou H-FABP reagent kit and Holland H-FABP kit.Correlation of the two kits was evaluated and Kappa test was used to examine the consistency of the results of the twoproducts. Results The sensitivity of H-FABP diagnosis of ACS detection Lanzhou kit was 91.8%, the specificity was 88.7%, andthe total diagnostic rate was 90.42% . The sensitivity of H-FABP diagnosis of ACS detection Holland kit was 90.3% , thespecificity was 86.8%, and the total diagnostic rate was 88.75%. The test results showed that two products yielded comparableresults (P=0.668, >0.05) with a good consistency (Kappa=0.726, P<0.01). Conclusions H-FABP ELISA detection kit producctedby LanZhou biological research institute has a good correlation with H-FABP detection kit produced by HBT company ofHolland and has the potential for clinical application.
0 / / Recommend
Add to citation manager EndNote|Ris|BibTeX
URL: https://www.j-smu.com/EN/
https://www.j-smu.com/EN/Y2014/V34/I06/901