Journal of Southern Medical University ›› 2014, Vol. 34 ›› Issue (05): 683-.

Previous Articles     Next Articles

Methodological assessment and bias estimation of two serum procalcitonin analysis
systems

  

  • Online:2014-05-20 Published:2014-05-20

Abstract: Objective To evaluate the comparability and bias of the test results of two detection systems for serum procalcitonin
(PCT) under the same laboratory condition.Methods According to the profile NCCLS-EP9-A, the two systems were used to
detect PCT to obtain the correlation coefficient and the liner equation for evaluation of the test result bias. Results and
Conclusion The test results of PCT showed no significant difference between the two detection systems (P>005) with a kappa
value greater than 0.75.The correlation coefficients of both systems were above 0.975, suggesting a consistency between them
for clinical detection of PCT.