Journal of Southern Medical University ›› 2014, Vol. 34 ›› Issue (05): 683-.
Previous Articles Next Articles
Online:
Published:
Abstract: Objective To evaluate the comparability and bias of the test results of two detection systems for serum procalcitonin(PCT) under the same laboratory condition.Methods According to the profile NCCLS-EP9-A, the two systems were used todetect PCT to obtain the correlation coefficient and the liner equation for evaluation of the test result bias. Results andConclusion The test results of PCT showed no significant difference between the two detection systems (P>005) with a kappavalue greater than 0.75.The correlation coefficients of both systems were above 0.975, suggesting a consistency between themfor clinical detection of PCT.